News | July 20, 2008

Dual-Source CT Lowers Dose on Heart Scan

July 17, 2008 – A new study reveals that, with dual-source computed tomography (DSCT), the dosage for a heart examination can be significantly lowered, in comparison to conventional computed tomography.

The study also demonstrated that stenoses can be diagnosed with similar high accuracy as with invasive X-ray angiography. At the University Hospital Zürich in Switzerland, 120 patients with suspected coronary heart disease were scanned with the world’s first CT scanner with two X-ray tubes, a SOMATOM Definition from Siemens Healthcare. The Siemens application, Adaptive Cardio Sequence with the step-and-shoot mode, was also used for the first time with a dual-source CT. The results of the study were published in the June issue of Heart, the official journal of the British Cardiovascular Society.

Researchers at the university hospital set out to see whether the dose used during CT angiography can be reduced. The research team used Adaptive Cardio Sequence, which was recently developed by Siemens for all SOMATOM Definition Scanners. The application is based on the step-and-shoot method, in which the next respective diastole is calculated and only this phase of the cardiac cycle is used to obtain image data. The combination of Adaptive Cardio Sequence and the DSCT’s maximum time resolution of 83 msec reportedly facilitates this technique and permits significant reduction in dosage.

“The results show that CT coronary angiography with a dual-source CT in step-and-shoot mode produces images of excellent diagnostic quality in patients with stable heart rates up to 70 bpm. The dosage reduction achieved in comparison to previous CT angiography is also remarkable. In our study, we required an effective dose of 2.5 mSv on average with a deviation of plus/minus 0.8 mSv. In the literature, a normal average effective dose for heart scans of between 9 mSv and 21 mSv is reported,” said Hatem Alkadhi, M.D., specialist in radiology at the Institute for Diagnostic Radiology, University Hospital Zürich.

The Adaptive Cardio Sequence supplements the step-and-shoot method with algorithms that monitor the heart frequency of patients and, if necessary, respond to arrhythmia, such as extrasystoles, additional heart beats. In this case, the recording phase is automatically postponed in order to avoid image errors due to the sudden movement. In addition, the recording window can be widened a little more if required so that, with the Adaptive Cardio Sequence, the robustness of the CT scan can be increased significantly overall.

H. Scheffel, MD, H. Alkadhi, MD, S. Leschka, MD, A. Plass, L. Desbiolles, I. Gruber..et al: Low-Dose CT Coronary Angiography in the Step-and-Shoot Mode: Diagnostic Performance, in Heart online, June 2, 2008.

For more information: http://www.siemens.com/healthcare

Related Content

iSchemaView Brings RAPID Imaging Platform to Australia and New Zealand
News | Stroke | July 13, 2018
iSchemaView has signed Diagnostic Imaging Australia (DIA) to be the exclusive distributor for the RAPID cerebrovascular...
HeartFlow Announces New Commercial Coverage With UnitedHealthcare
News | Computed Tomography (CT) | July 03, 2018
HeartFlow Inc. announced that UnitedHealthcare now covers the HeartFlow FFRct Analysis, extending access to their 45...
Post-Mortem CT Angiography Illuminates Causes of Death
News | CT Angiography (CTA) | June 25, 2018
Computed tomography (CT) angiography is a useful adjunct to autopsy that is likely to increase the quality of post-...
360 Photos | 360 View Photos | June 14, 2018
This is a 360 degree image from the Canon Aquilion 64-slice...
360 Photos | 360 View Photos | May 02, 2018
This is a 360 degree image from the Canon Aquilion One 320-slice computed tomography (CT) system installed at...
iSchemaView Receives FDA Clearance for Rapid CTA
Technology | CT Angiography (CTA) | May 01, 2018
Cerebrovascular imaging analysis company iSchemaView received final clearance from the U.S. Food and Drug...
HeartFlow FFRct Analysis Added to ITP Payment Program in the U.K.
News | CT Angiography (CTA) | April 25, 2018
HeartFlow Inc. announced that the National Health Service (NHS) England has chosen the HeartFlow FFRct (fractional flow...
New Policy Decisions Give Millions Access to HeartFlow FFRct Analysis
News | Cardiac Imaging | February 01, 2018
February 1, 2018 – HeartFlow announced that seven new commercial payers issued positive medical policies covering the
Overlay Init