News | Procedure Navigation Systems | February 05, 2019

Medtronic Recalls Synergy Cranial Software and Stealth Station S7 Cranial Software

Class I recall initiated due to inaccuracies displayed during surgical procedures

Medtronic Recalls Synergy Cranial Software and Stealth Station S7 Cranial Software

February 5, 2019 — Medtronic is recalling the Synergy Cranial Software and StealthStation S7 Cranial Software used with the StealthStation Surgical Navigation System due to reports of incorrect information displaying during biopsy procedures. The company said this could result in serious or life-threatening patient harm.

The Synergy Cranial Software and StealthStation S7 Cranial Software, used with the StealthStation Surgical Navigation System, provides detailed three-dimensional (3-D) images of a patient's brain to help neurosurgeons safely navigate surgical tools and implants used during brain surgery.

During a biopsy procedure, Medtronic said the software monitor may show that the tip of the surgical tool has not yet reached the planned target and may prevent the neurosurgeon from being able to accurately see the location of surgical tools in the patient's brain. In the event this software defect occurs, the neurosurgeon could potentially insert the surgical tool too deeply and damage the patient's healthy tissue, brain or blood vessels.

On Sept. 21, 2018, Medtronic sent an Urgent Medical Device Correction notice to customers. The notice provided the following mitigation to eliminate the risk of damaging healthy tissue:

  • Do not set a projection longer than the length of the surgical plan;
  • Follow the Cranial Software Pocket Guide, and always use the Biopsy Needle Mechanical Depth Stop; and
  • Ensure the Navigate Instrument Tip setting is selected prior to locking the trajectory and navigating the instrument.

The notice also instructed customers to take the following actions:

  • Review the notice and ensure appropriate staff are aware of the issue;
  • Sign and date the Consignee Response Form attached to the notice, and return the form to Medtronic by email at [email protected] or fax to Medtronic Technical Services at 651-367-7075 within 30 days of receipt.

On Dec. 28, Medtronic reported they were continuing to follow up with customers about  their Urgent Medical Device Correction over the next months.

Affected model numbers include:

  • 9733763 with software versions 2.2.0, 2.2.5, 2.2.6, 2.2.7
  • 9735585 with software versions 3.0.0, 3.0.1, 3.0.2, 3.1.0

All affected devices were manufactured from Aug. 31, 2011, to Oct. 16, 2018, and were distributed between the same dates.

Customers who have questions or need additional information or support related to this recall should contact Medtronic Technical Services at 1-888-826-5603. Healthcare professionals and consumers may report adverse reactions or quality problems they experience using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by fax to 1-800-FDA-0178.

For more information: www.medtronic.com

Related Content

Sectra Adds DePuy Synthes 3-D Templates to Pre-Operative Orthopedic Solution
News | Orthopedic Imaging | October 17, 2019
International medical imaging information technology (IT) and cybersecurity company Sectra is extending its pre-...
Guerbet Signs Agreement With Icometrix for Exclusive Distribution of Icobrain
News | Neuro Imaging | October 16, 2019
Guerbet announced it has signed an exclusive agreement with Icometrix for the distribution in France, Italy and Brazil...
RSNA Announces Intracranial Hemorrhage AI Challenge
News | Artificial Intelligence | October 08, 2019
The Radiological Society of North America (RSNA) recently launched its third annual artificial intelligence (AI)...
Brainlab Introduces Loop-X Mobile Intraoperative Imaging Robot
News | Mobile C-Arms | September 26, 2019
Digital medical technology company Brainlab unveiled Loop-X, which it calls the first mobile intraoperative imaging...
Virtual Reality 3-D Models Help Yield Better Surgical Outcomes

Joseph Shirk, M.D., of UCLA with the virtual reality headset. Image courtesy of UCLA Jonsson Comprehensive Cancer Center

News | Virtual and Augmented Reality | September 25, 2019
A UCLA-led study has found that using three-dimensional virtual reality (VR) models to prepare for kidney tumor...
Imaging Biometrics and Medical College of Wisconsin Awarded NIH Grant
News | Neuro Imaging | September 09, 2019
Imaging Biometrics LLC (IB), in collaboration with the Medical College of Wisconsin (MCW), has received a $2.75 million...
Neurological Brain Markers Might Detect Risk for Psychotic Disorders

Researchers at the University of Missouri used MRI scans similar to this photo to find neurological markers in the human brain. These markers can be used to detect people at risk for developing psychotic disorders and to understand when this risk has been successfully treated. Image courtesy of Marquette University/John Kerns.

News | Neuro Imaging | September 04, 2019
Help may be on the way for people who might lose contact with reality through a psychotic disorder, such as...
Delaware Imaging Network Now Offers NeuroQuant Brain Imaging MRI Software
News | Neuro Imaging | August 29, 2019
Delaware Imaging Network (DIN), Delaware’s largest network of outpatient medical imaging centers, has added NeuroQuant...
Smoldering Spots in the Brain May Signal Severe MS

NIH researchers found that dark rimmed spots representing ongoing, “smoldering” inflammation, may be a hallmark of more disabling forms of multiple sclerosis. Image courtesy of Reich lab, NIH/NINDS.

News | Neuro Imaging | August 22, 2019
Aided by a high-powered brain scanner and a 3-D printer, National Institutes of Health (NIH) researchers peered inside...
New MRI Technique Captures Brain Changes in Near-real Time

Differences in stiffness between stimulus states. Image courtesy of Patz et al.

News | Neuro Imaging | August 19, 2019
An international team of researchers developed a new magnetic resonance imaging (MRI) technique that can capture an...