News | Procedure Navigation Systems | February 05, 2019

Medtronic Recalls Synergy Cranial Software and Stealth Station S7 Cranial Software

Class I recall initiated due to inaccuracies displayed during surgical procedures

Medtronic Recalls Synergy Cranial Software and Stealth Station S7 Cranial Software

February 5, 2019 — Medtronic is recalling the Synergy Cranial Software and StealthStation S7 Cranial Software used with the StealthStation Surgical Navigation System due to reports of incorrect information displaying during biopsy procedures. The company said this could result in serious or life-threatening patient harm.

The Synergy Cranial Software and StealthStation S7 Cranial Software, used with the StealthStation Surgical Navigation System, provides detailed three-dimensional (3-D) images of a patient's brain to help neurosurgeons safely navigate surgical tools and implants used during brain surgery.

During a biopsy procedure, Medtronic said the software monitor may show that the tip of the surgical tool has not yet reached the planned target and may prevent the neurosurgeon from being able to accurately see the location of surgical tools in the patient's brain. In the event this software defect occurs, the neurosurgeon could potentially insert the surgical tool too deeply and damage the patient's healthy tissue, brain or blood vessels.

On Sept. 21, 2018, Medtronic sent an Urgent Medical Device Correction notice to customers. The notice provided the following mitigation to eliminate the risk of damaging healthy tissue:

  • Do not set a projection longer than the length of the surgical plan;
  • Follow the Cranial Software Pocket Guide, and always use the Biopsy Needle Mechanical Depth Stop; and
  • Ensure the Navigate Instrument Tip setting is selected prior to locking the trajectory and navigating the instrument.

The notice also instructed customers to take the following actions:

  • Review the notice and ensure appropriate staff are aware of the issue;
  • Sign and date the Consignee Response Form attached to the notice, and return the form to Medtronic by email at [email protected] or fax to Medtronic Technical Services at 651-367-7075 within 30 days of receipt.

On Dec. 28, Medtronic reported they were continuing to follow up with customers about  their Urgent Medical Device Correction over the next months.

Affected model numbers include:

  • 9733763 with software versions 2.2.0, 2.2.5, 2.2.6, 2.2.7
  • 9735585 with software versions 3.0.0, 3.0.1, 3.0.2, 3.1.0

All affected devices were manufactured from Aug. 31, 2011, to Oct. 16, 2018, and were distributed between the same dates.

Customers who have questions or need additional information or support related to this recall should contact Medtronic Technical Services at 1-888-826-5603. Healthcare professionals and consumers may report adverse reactions or quality problems they experience using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail or by fax to 1-800-FDA-0178.

For more information: www.medtronic.com

Related Content

New Study Evaluates Head CT Examinations and Patient Complexity
News | Neuro Imaging | May 17, 2019
Computed tomography (CT) of the head uses special X-ray equipment to help assess head injuries, dizziness and other...
Brain images that have been pre-reviewed by the Viz.AI artificial intelligence software to identify a stroke. The software automatically sends and alert to the attending physician's smartphone with links to the imaging for a final human assessment to help speed the time to diagnosis and treatment. Depending on the type of stroke, quick action is needed to either activate the neuro-interventional lab or to administer tPA. Photo by Dave Fornell.

Brain images that have been pre-reviewed by the Viz.AI artificial intelligence software to identify a stroke. The software automatically sends and alert to the attending physician's smartphone with links to the imaging for a final human assessment to help speed the time to diagnosis and treatment. Depending on the type of stroke, quick action is needed to either activate the neuro-interventional lab or to administer tPA. Photo by Dave Fornell.

Feature | Artificial Intelligence | May 17, 2019 | Inga Shugalo
With its increasing role in medical imaging,...
Videos | Advanced Visualization | May 16, 2019
This is an example of how virtual reality is being used in neuro-radiology to better evaluate patients using advanced
360 degree view inside an Interventional radiology lab at Henry Ford Hospital used for neuro-interventions and stroke.
360 Photos | 360 View Photos | May 06, 2019
This is a 360 degree photo view inside the neuro-interventional radiology lab at ...
California Hospital Adds Machine-Vision Image Guided Surgery Platform to New Operating Suites
News | Advanced Visualization | April 26, 2019
Pickup Family Neurosciences Institute at Hoag in Newport Beach, Calif., announced the addition of the 7D Surgical...
Gamma Knife radiosurgery has become the preferred radiation therapy option for patients with brain tumors at facilities like the Northwestern Medicine Cancer Center, pictured here

Gamma Knife radiosurgery has become the preferred radiation therapy option for patients with brain tumors at facilities like the Northwestern Medicine Cancer Center, pictured here. The technology is favored largely for its ability to precisely target tumors while sparing healthy tissue.

Feature | Radiation Oncology | April 11, 2019 | By Jeff Zagoudis
Brain tumors are some of the most complicated forms of cancer to treat due to their extremely sensitive location.
Videos | Artificial Intelligence | March 28, 2019
ITN Editor Dave Fornell takes a tour of some of the most interesting new...
NIH Study of Brain Energy Patterns Provides New Insights into Alcohol Effects

NIH scientists present a new method for combining measures of brain activity (left) and glucose consumption (right) to study regional specialization and to better understand the effects of alcohol on the human brain. Image courtesy of Ehsan Shokri-Kojori, Ph.D., of NIAAA.

News | Neuro Imaging | March 22, 2019
March 22, 2019 — Assessing the patterns of energy use and neuronal activity simultaneously in the human brain improve
Book Chapter Reports on Fonar Upright MRI for Hydrocephalus Imaging

Rotary misalignment of atlas (C1) and axis (C2). Image courtesy of Scott Rosa, DC, BCAO.

News | Magnetic Resonance Imaging (MRI) | March 20, 2019
Fonar Corp. reported publication of a chapter where the physician-author-researchers utilized the Fonar Upright Multi-...
Gregory W. Albers, M.D., Receives Distinguished Clinical Research Achievement Award for Stroke Research
News | Stroke | March 19, 2019
iSchemaView announced the company’s co-founder Gregory Albers, M.D., has received the Distinguished Clinical Research...