Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.

23-Year-Old Man with Body Mass Index of 25.1. Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.


June 7, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), ultrasound-derived fat fraction (UDFF) is strongly associated with MRI proton-density fat fraction (PDFF) and provides high sensitivity for detecting hepatic steatosis.

Noting that a reduced number of measurements is sufficient for determining overall UDFF (i.e., median value of the median of the five measurements for each of three acquisitions), “a UDFF cutoff of >5% provides high AUC and sensitivity, albeit low specificity, for detection of MRI PDFF ≥5.5%,” clarified corresponding author Andrew T. Trout of Cincinnati Children’s Hospital Medical Center in Ohio.

Trout and team’s cross-sectional study registered 56 overweight and obese adolescents and adults (≥16 years) who underwent investigational ultrasound (ACUSON Sequoia; deep abdominal transducer) and MRI examinations of the liver during a single visit from August to October 2020. Ultrasound examinations included three UDFF acquisitions of five measurements each. MRI examinations included three PDFF acquisitions with calculation of an overall median PDFF.

Ultimately, UDFF was positively associated with MRI PDFF (rho=0.82 [95% 0.71-0.89]; p<.001), with mean bias between methods of 4.0%. In addition, UDFF had an AUC of 0.90 (95% CI: 0.79-0.96, p<.001) for identifying an MRI PDFF ≥5.5%, with a cutoff value of >5% having 94.1% sensitivity and 63.6% specificity.

“In clinical practice, the authors of this AJR article concluded, “three measurements are likely sufficient for determining UDFF.”

For more information: www.arrs.org


Related Content

News | X-Ray

March 20, 2023 — Researchers from UNSW Sydney have developed an algorithm which produces high-resolution modeled images ...

Time March 20, 2023
arrow
News | Artificial Intelligence

March 17, 2023 — A new paper published in the Journal of Medical Internet Research describes how generative models such ...

Time March 17, 2023
arrow
Feature | Magnetic Resonance Imaging (MRI) | By Brad Sutton

March 16, 2023, marks 50 years since Paul Lauterbur published his seminal Nature paper establishing zeugmatography — now ...

Time March 16, 2023
arrow
News | Lung Imaging

March 15, 2023 — According to an accepted manuscript published in ARRS’ own American Journal of Roentgenology (AJR) ...

Time March 15, 2023
arrow
News | Coronavirus (COVID-19)

March 13, 2023 — According to an accepted manuscript published in ARRS’ American Journal of Roentgenology (AJR) ...

Time March 13, 2023
arrow
News | Artificial Intelligence

March 10, 2023 — Breast cancer undisputedly has the highest incidence rate in female patients. Moreover, out of the six ...

Time March 10, 2023
arrow
News | Contrast Media

March 9, 2023 — Contrast agents used in MRI scans to improve the clarity of imaging for diagnostic accuracy, accumulate ...

Time March 09, 2023
arrow
Feature | Ultrasound Imaging | By Robert Bard, MD, DABR, FAIUM, FASLMS

Electromagnetic therapies currently used in the National Aeronautics and Space Administration (NASA) space centers ...

Time March 08, 2023
arrow
Feature | Ultrasound Imaging | By Christine Book

Compared to a marked slowdown of demand during the pandemic, the ultrasound systems market is making up for lost time ...

Time March 08, 2023
arrow
News | Contrast Media

March 2, 2023 — GE HealthCare announced at the European Congress of Radiology (ECR) in Vienna, Austria, the completion ...

Time March 02, 2023
arrow
Subscribe Now