Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.

23-Year-Old Man with Body Mass Index of 25.1. Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.


June 7, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), ultrasound-derived fat fraction (UDFF) is strongly associated with MRI proton-density fat fraction (PDFF) and provides high sensitivity for detecting hepatic steatosis.

Noting that a reduced number of measurements is sufficient for determining overall UDFF (i.e., median value of the median of the five measurements for each of three acquisitions), “a UDFF cutoff of >5% provides high AUC and sensitivity, albeit low specificity, for detection of MRI PDFF ≥5.5%,” clarified corresponding author Andrew T. Trout of Cincinnati Children’s Hospital Medical Center in Ohio.

Trout and team’s cross-sectional study registered 56 overweight and obese adolescents and adults (≥16 years) who underwent investigational ultrasound (ACUSON Sequoia; deep abdominal transducer) and MRI examinations of the liver during a single visit from August to October 2020. Ultrasound examinations included three UDFF acquisitions of five measurements each. MRI examinations included three PDFF acquisitions with calculation of an overall median PDFF.

Ultimately, UDFF was positively associated with MRI PDFF (rho=0.82 [95% 0.71-0.89]; p<.001), with mean bias between methods of 4.0%. In addition, UDFF had an AUC of 0.90 (95% CI: 0.79-0.96, p<.001) for identifying an MRI PDFF ≥5.5%, with a cutoff value of >5% having 94.1% sensitivity and 63.6% specificity.

“In clinical practice, the authors of this AJR article concluded, “three measurements are likely sufficient for determining UDFF.”

For more information: www.arrs.org


Related Content

News | PET Imaging

April 24, 2024 — A new study from Brigham and Women’s Hospital, a founding member of the Mass General Brigham healthcare ...

Time April 24, 2024
arrow
News | Contrast Media

April 24, 2024 — The International Contrast Ultrasound Society (ICUS) and Northwest Imaging Forums (NWIF) announced an ...

Time April 24, 2024
arrow
News | Radiology Business

April 23, 2024 — A diverse writing group—lead by authors at the University of Toronto—have developed an approach for ...

Time April 23, 2024
arrow
News | Radiology Business

April 17, 2024 — VISTA.AI announced the appointment of Daniel Hawkins as President and CEO. The company is pioneering AI ...

Time April 17, 2024
arrow
News | Magnetic Resonance Imaging (MRI)

April 8, 2024 — Magnetic resonance-guided focused ultrasound (MRgFUS) is a non-invasive technique for neuroregulation ...

Time April 08, 2024
arrow
News | Magnetic Resonance Imaging (MRI)

April 5, 2024 — Osteoarthritis — not age — may play the greatest role in determining the utility of MRI for patients 45 ...

Time April 05, 2024
arrow
News | Radiology Business

April 4, 2024 — FUJIFILM Healthcare Americas Corporation, a leading provider of diagnostic and enterprise imaging ...

Time April 04, 2024
arrow
News | Radiation Oncology

April 2, 2024 — In a 10-center study, microwave ablation offered progression free survival rates and fewer complications ...

Time April 02, 2024
arrow
News | Molecular Imaging

March 29, 2024 — Magnetic resonance imaging (MRI) is a cornerstone in the landscape of medical diagnostics, celebrated ...

Time March 29, 2024
arrow
News | Electronic Medical Records (EMR)

March 28, 2024 — Immediately releasing radiology reports under the 21st Century Cures Act (Cures Act) empowers patients ...

Time March 28, 2024
arrow
Subscribe Now