Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.

23-Year-Old Man with Body Mass Index of 25.1. Left: UDFF—operator places crossbar at liver capsule, with sample ROI fixed 1.5 cm deep from crossbar, to ensure measurement obtained sufficiently deep to liver capsule. Overall UDFF was 3%. Right: Median MRI PDFF from three acquisitions was 3%, demonstrating agreement.


June 7, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), ultrasound-derived fat fraction (UDFF) is strongly associated with MRI proton-density fat fraction (PDFF) and provides high sensitivity for detecting hepatic steatosis.

Noting that a reduced number of measurements is sufficient for determining overall UDFF (i.e., median value of the median of the five measurements for each of three acquisitions), “a UDFF cutoff of >5% provides high AUC and sensitivity, albeit low specificity, for detection of MRI PDFF ≥5.5%,” clarified corresponding author Andrew T. Trout of Cincinnati Children’s Hospital Medical Center in Ohio.

Trout and team’s cross-sectional study registered 56 overweight and obese adolescents and adults (≥16 years) who underwent investigational ultrasound (ACUSON Sequoia; deep abdominal transducer) and MRI examinations of the liver during a single visit from August to October 2020. Ultrasound examinations included three UDFF acquisitions of five measurements each. MRI examinations included three PDFF acquisitions with calculation of an overall median PDFF.

Ultimately, UDFF was positively associated with MRI PDFF (rho=0.82 [95% 0.71-0.89]; p<.001), with mean bias between methods of 4.0%. In addition, UDFF had an AUC of 0.90 (95% CI: 0.79-0.96, p<.001) for identifying an MRI PDFF ≥5.5%, with a cutoff value of >5% having 94.1% sensitivity and 63.6% specificity.

“In clinical practice, the authors of this AJR article concluded, “three measurements are likely sufficient for determining UDFF.”

For more information: www.arrs.org

Related Content

News | Radiology Business

July 1, 2022 — Richard Fabian has been promoted to the role of Chief Executive Officer (CEO) of Fujifilm Sonosite, Inc ...

Time July 01, 2022
arrow
News | Prostate Cancer

July 1, 2022 — Insightec, a global healthcare company dedicated to using acoustic energy to transform patient care ...

Time July 01, 2022
arrow
News | Computed Tomography (CT)

June 30, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), a 5-tiered CT scoring algorithm may ...

Time June 30, 2022
arrow
News | Information Technology

June 29, 2022 — Bayer unveiled Calantic Digital Solutions, a new cloud-hosted platform delivering access to digital ...

Time June 29, 2022
arrow
News | Coronavirus (COVID-19)

June 27, 2022 — Three articles and an accompanying editorial provide information on the effects of Long COVID in the ...

Time June 28, 2022
arrow
News | MRI Breast

June 22, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), contrast-enhanced mammography (CEM) may be a ...

Time June 22, 2022
arrow
News | Artificial Intelligence

June 14, 2022 — According to ARRS’ American Journal of Roentgenology (AJR), incorporating AI support into clinical ...

Time June 14, 2022
arrow
News | Artificial Intelligence

June 9, 2022 — Infinitt North America and Brainreader A/S announced today a wide ranging and unique partnership to ...

Time June 09, 2022
arrow
Feature | Digital Radiography (DR) | By ResearchAndMarkets

The "Mobile Imaging Services - Global Market Trajectory & Analytics" report has been added to ResearchAndMarkets.com's ...

Time June 09, 2022
arrow
News | Radiation Therapy

June 3, 2022 — Henry Ford Health is the first in the world to complete a full course of patient treatments using the ...

Time June 03, 2022
arrow
Subscribe Now