April 14, 2011 – A class I recall has been issued for the Xoft Axxent Flexishield Mini because the product may shed particles of tungsten that can be mistaken for suspicious breast calcifications during follow up X-rays or scans.

The device consists of a circular silicone rubber pad containing tungsten particulate. The pad is 12.7 cm in diameter and 0.1 cm (1 mm) thick. It is designed to shape a therapeutic radiation therapy beam from a low energy source to provide a means to accommodate different shapes of treatment fields. It also limits the amount of radiation encountered by the skin or healthy tissue by shielding areas that do not need radiation therapy.

The pad should be placed over the surface that needs shielding from the radiation therapy. It can be cut to accommodate the shape of the radiation therapy beam. It can be used on the skin, and it can be used to shield internal tissue from radiation therapy during intraoperative radiation therapy (IORT).

This product was manufactured from Sept. 11, 2009 through June 28, 2010, and distributed nationwide from Oct. 2009 through Dec. 2010.

Xoft conducted an extensive and comprehensive failure investigation and risk assessment regarding this matter. After laboratory testing and a comprehensive literature review, Xoft has concluded that there is no evidence that these tungsten particles are toxic – only a few health effects have been reported in humans with this material, each of which involved exposure at much higher levels. No permanent impairment of bodily functions or permanent damage to body structures is anticipated.

On Feb. 3, 2011, the company sent an Urgent Field Removal Action letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.

Customers were instructed to stop using all units of the Flexishield Mini Catalog Number F5300 in their inventory and return them to the company.

Class 1 recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of these products will cause serious adverse health consequences or death.

For more information: (877) 963-8327


Related Content

News | Radiation Dose Management

April 25, 2024 — BIOTRONIK, a leading global medical technology company specializing in innovative cardiovascular and ...

Time April 25, 2024
arrow
News | Breast Imaging

April 2, 2024 — iCAD, Inc., a global leader in clinically proven AI-powered cancer detection solutions, announced ...

Time April 02, 2024
arrow
News | Artificial Intelligence

March 28, 2024 — As artificial intelligence (AI) makes its way into cancer care – and into discussions between ...

Time March 28, 2024
arrow
News | Prostate Cancer

March 27, 2024 — A minimally invasive treatment using MRI and transurethral ultrasound instead of surgery or radiation ...

Time March 27, 2024
arrow
News | Breast Imaging

March 20, 2024 — Interval breast cancer is one that appears between two reviews of a screening program. That is, it is ...

Time March 20, 2024
arrow
Videos | Radiation Oncology

In the conclusion of this 3-part video series on recent advancements in diagnostic radiology, current editorial advisory ...

Time March 19, 2024
arrow
News | Breast Imaging

March 18, 2024 — QT Imaging Holdings, Inc., a medical device company engaged in research, development, and ...

Time March 18, 2024
arrow
News | Breast Imaging

March 15, 2024 — Screening-detected radial scars without atypia at core-needle biopsy (CNB) have a low upstaging rate to ...

Time March 15, 2024
arrow
News | Breast Imaging

March 6, 2024 — There is a pressing need to explore and understand which social determinants of health (SDOH) and health ...

Time March 06, 2024
arrow
Feature | Radiation Oncology | By Christine Book

Appreciating the considerable advances in the clinical application of artificial intelligence (AI) within healthcare ...

Time March 06, 2024
arrow
Subscribe Now