Technology | July 26, 2011 | Dave Fornell

FDA: Stop Using CardioGen-82 Due to Increased Radiation Exposure

Patients set off U.S. border radiation detectors

July 26, 2011 – The U.S. Food and Drug Administration (FDA) is alerting healthcare professionals to stop using CardioGen-82 for cardiac positron emission tomography (PET) scans. The manufacturer, Bracco Diagnostics Inc., has decided to voluntarily recall CardioGen-82 this week after two patients set off radiation detectors at a U.S. border crossings due to strontium contamination.

It is the only FDA-approved generator-based PET myocardial perfusion agent. It uses a strontium (Sr)-82 generator to produce a rubidium (Rb)-82 radiotracer injection.

The FDA has received reports of two patients who underwent PET myocardial perfusion imaging scans with CardioGen-82 who had detectable levels of radiation several months after their PET scans. Both patients were crossing the border to/from the United States when radiation detectors identified radiation originating from the patients.

The radiation was found to originate from strontium (Sr-82 and Sr-85). A PET scan had been performed in each patient approximately two and four months earlier, respectively. Since Rb-82 has a half-life of 75 seconds and both patients reported that they had no other exposure to radioactive substances, the residual radiation is likely due to undetected strontium breakthrough.

The strontium isotope exposure may have originated from the myocardial perfusion imaging, said Alberto Spinazzi, M.D., senior vice president of global medical and regulatory affairs, Bracco Diagnostics, in a prepared statement from the company.  He said neither strontium-exposed patient suffered any acute health effects or required treatment. However, the long-term effects, if any, are not known.

The radiation exposure that a patient would typically receive from an injection of Rb-82 chloride (approximately 2.8 mSv) is generally less than that associated with other radiotracer cardiac diagnostic scans.  The radiation exposure that the two patients received due to strontium isotope exposure appears to substantially exceed that typically associated with CardioGen-82.  This assessment is based upon modeling performed by the Los Alamos National Laboratory.  This modeling suggests that the excessive radiation exposure (approximately 90 mSv) associated with the strontium isotopes appears similar to the amount of cumulative radiation exposure some patients receive during cardiac diagnostic evaluations with other radionuclides. 

Based on further investigation, the FDA determined that the current CardioGen-82 manufacturing procedures are not sufficient to ensure reliable performance of the generator used to produce the Rb-82 chloride injections. The FDA stated reliable generator performance is essential to help prevent strontium breakthrough from CardioGen-82 and to prevent patients from being exposed to excess radiation. FDA is also currently investigating the sufficiency of the testing procedures used to detect strontium breakthrough at the clinical sites that use CardioGen-82. 

The FDA recommends that healthcare professionals use alternatives to the CardioGen-82 generator when planning nuclear medicine cardiac scans. Patients who have any questions or concerns should talk to their healthcare professional. 

The FDA is continuing to work with the Nuclear Regulatory Commission and Bracco to determine the root cause for the increased radiation exposure detected in the two patients. The extent to which any additional patients may have received inadvertent radiation exposure is also under investigation. FDA plans to notify the public with updates.

The total number of patients who have received doses from CardioGen-82 generators with strontium breakthrough is currently unknown. CardioGen-82 was first cleared by the FDA in 1989.

Bracco said the strontium breakthrough can occur with CardioGen-82 and necessitates close attention to mandatory daily quality control testing to help minimize radiation risks to patients. The company is emphasizing to all CardioGen-82 users the importance of strict adherence to the instructions contained in the prescribing information for CardioGen-82 regarding mandatory daily strontium breakthrough testing (section 2.7 of the full prescribing information). The user training material for CardioGen-82 says testing for strontium breakthrough may sometimes need to be performed twice daily.

For more information: www.fda.gov/Drugs/DrugSafety/ucm265278.htm

 

 

Related Content

CDC and FDA Call for Pause on Janssen COVID-19 Vaccine Due to Rare Blood Clots #COVID19 #Janssen
Feature | Coronavirus (COVID-19) | April 15, 2021 | By Dave Fornell
April 15, 2021 — The U.S.
ScImage, Inc. celebrates its cloud partnership with Digirad Health (“Digirad”), a division of Star Equity Holdings, after a year of successful deployment of PICOM365 for mobile imaging. Digirad’s fleet of mobile SPECT, echocardiology, vascular and general ultrasound units combined with PICOM365’s cloud image management workflow leverage each company’s strengths to create an exemplary reading and reporting environment.
News | Archive Cloud Storage | March 30, 2021
March 30, 2021 — ScImage, Inc.

Scan of brain of someone who did not develop TTS. Image courtesy of European Heart Journal

News | PET-CT | March 29, 2021
March 29, 2021 — Heightened activity in the brain, caused by stressful events, is linked to the risk of developing a
Shine executives are joined by the company’s construction managers and partners at its medical isotope production facility. The group was commemorating the facility’s achievement of weathertight status

Shine executives are joined by the company’s construction managers and partners at its medical isotope production facility. The group was commemorating the facility’s achievement of weathertight status. Image courtesy of Shine Medical Technologies

News | Radiopharmaceuticals and Tracers | March 26, 2021
March 26, 2021 — Shine Medical Technologies LLC announced that construction of the exterior structure of its first-of
Neurologica Mobile Stroke Unit includes new CT scanner and features that help EMTs improve patient care
News | Computed Tomography (CT) | March 17, 2021
March 17, 2021 — NeuroLogica Corp.
The RapidAI footprint grew to over 60 countries, with its technology utilized in over 1,800 hospitals
News | Artificial Intelligence | March 16, 2021
March 16, 2021 — RapidAI, a leader in advanced...
The phantom used for hyperpolarized imaging, with an illustration of imaging slices acquired using the new technique. photo/©: Laurynas Dagys, University of Southampton

The phantom used for hyperpolarized imaging, with an illustration of imaging slices acquired using the new technique. photo/©: Laurynas Dagys, University of Southampton

News | Nuclear Imaging | March 12, 2021
March 12, 2021 — Magnetic resonance imaging (MRI)
Diagnostic imaging studies using Tc-99m inform healthcare decisions for approximately 30 million patients annually around the world
News | Nuclear Imaging | March 08, 2021
March 8, 2021 — IBA (Ion Beam Applications S.A., EURONEXT) an