Technology | July 26, 2011 | Dave Fornell

FDA: Stop Using CardioGen-82 Due to Increased Radiation Exposure

Patients set off U.S. border radiation detectors

July 26, 2011 – The U.S. Food and Drug Administration (FDA) is alerting healthcare professionals to stop using CardioGen-82 for cardiac positron emission tomography (PET) scans. The manufacturer, Bracco Diagnostics Inc., has decided to voluntarily recall CardioGen-82 this week after two patients set off radiation detectors at a U.S. border crossings due to strontium contamination.

It is the only FDA-approved generator-based PET myocardial perfusion agent. It uses a strontium (Sr)-82 generator to produce a rubidium (Rb)-82 radiotracer injection.

The FDA has received reports of two patients who underwent PET myocardial perfusion imaging scans with CardioGen-82 who had detectable levels of radiation several months after their PET scans. Both patients were crossing the border to/from the United States when radiation detectors identified radiation originating from the patients.

The radiation was found to originate from strontium (Sr-82 and Sr-85). A PET scan had been performed in each patient approximately two and four months earlier, respectively. Since Rb-82 has a half-life of 75 seconds and both patients reported that they had no other exposure to radioactive substances, the residual radiation is likely due to undetected strontium breakthrough.

The strontium isotope exposure may have originated from the myocardial perfusion imaging, said Alberto Spinazzi, M.D., senior vice president of global medical and regulatory affairs, Bracco Diagnostics, in a prepared statement from the company.  He said neither strontium-exposed patient suffered any acute health effects or required treatment. However, the long-term effects, if any, are not known.

The radiation exposure that a patient would typically receive from an injection of Rb-82 chloride (approximately 2.8 mSv) is generally less than that associated with other radiotracer cardiac diagnostic scans.  The radiation exposure that the two patients received due to strontium isotope exposure appears to substantially exceed that typically associated with CardioGen-82.  This assessment is based upon modeling performed by the Los Alamos National Laboratory.  This modeling suggests that the excessive radiation exposure (approximately 90 mSv) associated with the strontium isotopes appears similar to the amount of cumulative radiation exposure some patients receive during cardiac diagnostic evaluations with other radionuclides. 

Based on further investigation, the FDA determined that the current CardioGen-82 manufacturing procedures are not sufficient to ensure reliable performance of the generator used to produce the Rb-82 chloride injections. The FDA stated reliable generator performance is essential to help prevent strontium breakthrough from CardioGen-82 and to prevent patients from being exposed to excess radiation. FDA is also currently investigating the sufficiency of the testing procedures used to detect strontium breakthrough at the clinical sites that use CardioGen-82. 

The FDA recommends that healthcare professionals use alternatives to the CardioGen-82 generator when planning nuclear medicine cardiac scans. Patients who have any questions or concerns should talk to their healthcare professional. 

The FDA is continuing to work with the Nuclear Regulatory Commission and Bracco to determine the root cause for the increased radiation exposure detected in the two patients. The extent to which any additional patients may have received inadvertent radiation exposure is also under investigation. FDA plans to notify the public with updates.

The total number of patients who have received doses from CardioGen-82 generators with strontium breakthrough is currently unknown. CardioGen-82 was first cleared by the FDA in 1989.

Bracco said the strontium breakthrough can occur with CardioGen-82 and necessitates close attention to mandatory daily quality control testing to help minimize radiation risks to patients. The company is emphasizing to all CardioGen-82 users the importance of strict adherence to the instructions contained in the prescribing information for CardioGen-82 regarding mandatory daily strontium breakthrough testing (section 2.7 of the full prescribing information). The user training material for CardioGen-82 says testing for strontium breakthrough may sometimes need to be performed twice daily.

For more information: www.fda.gov/Drugs/DrugSafety/ucm265278.htm

 

 

Related Content

EchoGo uses artificial intelligence (AI) to calculate cardiac ultrasound left ventricular ejection fraction (EF), the most frequently used measurement of heart function, left ventricular volumes (LV) and, for the first time for an AI application, automated cardiac strain.

EchoGo uses artificial intelligence (AI) to calculate cardiac ultrasound left ventricular ejection fraction (EF), the most frequently used measurement of heart function, left ventricular volumes (LV) and, for the first time for an AI application, automated cardiac strain.

News | Cardiovascular Ultrasound | November 14, 2019
November 14, 2019 — Ultromics has received 510(k) clearance from the U.S.
This chest X-ray of a patient being treated for e-cigarette or vaping-associated lung injury shows lung opacities, densities and whitish cloud-like areas which are typically seen with unusual pneumonias, fluid in lungs or lung inflammation. Image courtesy of Intermountain Healthcare

This chest X-ray of a patient being treated for e-cigarette or vaping-associated lung injury shows lung opacities, densities and whitish cloud-like areas which are typically seen with unusual pneumonias, fluid in lungs or lung inflammation. Image courtesy of Intermountain Healthcare

News | Clinical Trials | November 08, 2019
November 8, 2019 — As the outbreak of lung injuries and deaths associated with e-cigarettes, or...
Philips Medical System is recalling its older Forte Gamma Camera SPECT imaging systems due to the possibility of the detectors falling off of the unit onto the patient. The two gamma cameras can bee seen in this photo on either side of the patient bed. These can be rotated above the patient.

Philips Medical System is recalling its older Forte Gamma Camera SPECT imaging systems due to the possibility of the detectors falling off of the unit onto the patient. The two gamma cameras can be seen in this photo on either side of the patient bed. These can be rotated above the patient.

Feature | Nuclear Imaging | November 05, 2019 | Dave Fornell, Editor
November 5, 2019 — Philips Medical System is recalling the Forte Gamma Camera System due to the potential for the 660
 Phoenix’s fusion neutron generation technology.
News | Radiopharmaceuticals and Tracers | October 28, 2019
October 28, 2019 — Phoenix LLC and Shine Medical Technologies LLC, nuclear technology companies focused on near-term
The Konica Minolta Sonimage HS1 hand-carried system is designed for point of care ultrasound (POCUS) use for an immediate, quick look inside the patient. POCUS systems from several vendors are seeing rapid growth.

The Konica Minolta Sonimage HS1 hand-carried system is designed for point of care ultrasound (POCUS) use for an immediate, quick look inside the patient. POCUS systems from several vendors are seeing rapid growth.

Feature | Ultrasound Imaging | October 24, 2019 | Joan Toth
With the miniaturization of technology, improved ease of use, lower system cost, increased portability, and greater
Greater Left Ventricular Mass Increases Risk of Heart Failure
News | Cardiac Imaging | October 21, 2019
Elevated left ventricular mass, known as left-ventricular hypertrophy, is a stronger predictor of coronary artery...
USF Health Expands Digisonics System With Vascular Reporting
News | Cardiac PACS | October 17, 2019
University of South Florida (USF) Health in Tampa, Fla., has enhanced their use of the Digisonics Cardiovascular...
GE Healthcare and Theragnostics Partnering on PSMA PET/CT Imaging Agent
News | Prostate Cancer | October 16, 2019
GE Healthcare and Theragnostics have entered into a global commercial partnership for a new prostate-specific membrane...
FDA Clears Modules of AI-Rad Companion Chest CT From Siemens Healthineers
Technology | Computed Tomography (CT) | September 26, 2019
The U.S. Food and Drug Administration (FDA) has cleared three modules of AI-Rad Companion Chest CT, an intelligent...