Feature | Computed Tomography (CT) | August 27, 2010 | Dave Fornell

Costs vs. Benefits: Comparing 64-Slice to 256-, 320-Slice CT

Experts share thoughts on whether the diagnostic value justifies the cost of higher slice CT systems over standard 64-slice systems

cardiac CT, cardiovascular computed tomography, what to look for in a CT scanner, advances in CT

Users say the image quality from a Toshiba Aquilion One 320-slice scanner provides accurate diagnoses because the images do not have the artifacts common to stitched 64-slice images. A heart can be scanned in one volume during a single cardiac phase.

what to look for in a CT scanner, advances in CT

The GE Healthcare LightSpeed CT750 High Definition (HD) 64-slice CT system is much faster than standard 64-slice systems. It features Gemstone technology with adaptive statistical iterative reconstruction (ASIR).

In this era of financial cost cutting, questions are raised over the true need for 256- or 320-slice computed tomography (CT) systems for cardiac imaging. There are many radiologists and cardiologists who argue a 64-slice CT system is good enough to make a cardiac diagnosis without spending twice as much for the higher-slice systems. However, those who adopted 256- or 320-slice systems say the quality of the images is superior and leads to better diagnoses in CT angiography (CTA).

System users say a 64-slice scanner costs a little more than $1 million, while a 256-slice scanner runs about $2 million, and a 320-slice system costs about $2.5 million.
Imaging Technology News spoke with cardiac imaging specialists to gain their insights.

 

Watch an updated video from SCCT 2016 explaining the differences in newer generation CT systems — What to Consider When Comparing 64-slice to Higher Slice CT Systems. 

 

Watch video explaining recent trends in CT technology — “RSNA Technology Report 2015: Computed Tomography” — where Contributing Editor Greg Freiherr describes computed tomography advances shown at RSNA 2015

Better Diagnostic Images
The debate over the use of established and new technology happens with each new generation of CT, said Michael Vannier, M.D., FACR, professor of radiology and medicine at the University of Chicago Medical Center. “In an ideal world you have patients with low heart rates, they can hold their breath and they don’t have lots of disease. With these conditions you can get great images with just about any CT scanner,” he said.

However, the realities are some patients have arrhythmias, fast heart rates, are obese and pediatric patients squirm. He said these conditions are better suited for high-slice systems, which are faster and offer a larger imaging area. These were among the reasons why the University of Chicago purchased a Philips 256-slice Brilliance iCT system.

“The 256-slice scanner costs substantially more than the 64-slice systems. We were skeptical at first, but it was apparent from the start the quality of the exams was much better,” Vannier said. “I don’t think any of our physicians would use a 64-slice system any more for these cardiac examinations. We were accustomed in the past to having some studies that were just uninterpretable. But today, that has largely gone away.”

While the cost is high, the image quality from a Toshiba Aquilion One 320-slice scan provides much better patient care through more accurate diagnoses, said Michael Poon, M.D., FACC, professor of medicine and radiology, director of advanced cardiac imaging, Stony Brook University Medical Center, Stony Brook, N.Y.

The 320-slice volume scan offers 16 cm of imaging area in one rotation. This allows an image of the entire heart during one phase, rather than taking several images and stitching them together, as with 64-slice scanners. Poon explained 64-slice images generally have artifacts due to the multiple image stitching, where parts of the heart or vessels are misaligned due to cardiac movement. Table movement during 64-slice imaging can also contribute to artifacts. He said Siemens and other manufacturers use a continuous table movement. GE Healthcare’s scanners move in stages during scanning. The Aquilion One uses a stationary table because the imaging volume is so large.
“In 64-slice images, the top and the bottom come from different times, but ideally you want to have an image of the heart from the same time,” Poon said. “That is the main problem with cardiac CTA imaging. Ideally you need a scanner that images the whole heart at once. That’s why the 320-slice scanner opens up a whole new era in cardiac imaging.”

More Evidence is Needed
There are many cardiologists who insist 64-slice scanning is adequate for cardiac imaging and more clinical evidence is needed to justify the high-cost, higher-slice systems.

“I don’t see a clinical need for a 320- or 256-slice scanner right now,” said James Min, M.D., director of the cardiovascular CT lab at New York Presbyterian Hospital and assistant professor of radiology at Weill Cornell Medical College. “I think we are looking at a three- to five-year process before the incremental benefit justifies the increased cost.”

In recent years, there has been a lot of hype over the release of the 256- and 320-slice systems, but Min said hard, clinical evidence is needed before large numbers of institutions start making purchases. He has not yet seen any published studies of how 320-slice stacks up diagnostically to 64-slice systems.

“I think it’s not only the cart before the horse, but we don’t have a horse yet,” Min said. “We need to show 320-slice CT has a definite diagnostic benefit over 64-slice.”

A high-slice scanner also is not a cure-all for CT artifacts. He explained beam-hardening and some motion artifacts will still be seen with a 256- or 320-slice system.

Stress Testing on CT
Min said the higher-slice systems may offer an advantage in perfusion imaging over 64-slice, because of their ability to image the entire heart at once. Up until now, he said, only 64-slice systems have been evaluated for CT perfusion imaging. The higher-slice systems will likely show CT can be used for high-quality cardiac perfusion imaging, but clinical studies are needed. These systems also have additional perfusion applications for liver and brain imaging. This may help justify the higher cost, if shared between several specialities.

The Aquilion One 320-slice scanner can complete an entire body scan in three seconds, and a heart scan without table movement in less than half a second. Poon said this speed and coverage will allow CT adenosine stress imaging, which is primarily reserved only for magnetic resonance imaging (MRI) and nuclear imaging.

Changes in Reading Images
The slight misalignment artifacts in 64-slice cardiac images are taken into account when reading. However, a 320-slice scanner image eliminates these artifacts. Poon said what you see is the actual, full heart frozen at one particular time.

“I had to change my reading habits,” Poon said. “I was so used to reading 64-slice that I just assumed many things I saw were artifacts. But now what I am seeing is a lesion, it’s not an artifact any more. We are so used to artifacts on 64-slice that we discount many things. You have to look at the 320 under a different light.”

The University of Chicago uses 16-, 64- and 256-slice systems. Some patients had exams at different times performed by two or all three of the systems. Vannier said the radiologists compared these images and were able to pick out the 256 system immediately. All motion was frozen and lacked misalignment artifacts. The high motion area near the diaphragm also contributed to motion blur that appeared as a thickening of tissue, which would have led to an improper diagnosis. “We started asking ourselves if we were calling things abnormal when they were actually OK. It was really sobering,” Vannier explained.

Spacial Resolution vs. Slices
When New York Presbyterian Hospital recently decided to add a new scanner for its dedicated cardiac CT lab, increasing spacial resolution was the deciding factor over volume coverage.

“We had the option to go with any scanner, but what we are interested in is increased spacial resolution,” Min said. “But I don’t want spacial resolution at the cost of higher radiation dose.”

The hospital purchased a GE Healthcare LightSpeed CT750 High Definition (HD) 64-slice system. The scanner is supposed to produce images 100 times faster than standard 64-slice systems, with up to 47 percent greater image detail. It uses GE’s Gemstone technology with adaptive statistical iterative reconstruction (ASIR), offering improved spatial resolution that reduces calcium blooming artifacts. It also offers up to 50 percent less radiation dose without compromising image quality. The system also sees through calcium and stents better than previous 64-slice systems.

“It improved our diagnostic certainty and eliminated the need for additional tests,” Min said. This helped reduce the number of referrals for additional testing, which cut costs and the amount of time needed to assess a patient’s condition.

 

CT System Comparison Chart

View an apples-to-apples comparison of technical specifications on currently available computed tomography systems on the U.S. market at www.itnonline.com/content/computed-tomography-systems. To view the chart specs you will need to log-in, which is free and only takes a minute. 

Related Content

Fujifilm’s Sonosite SII POC ultrasound system helps to keep crowded areas clearer with a small ultrasound footprint.

Fujifilm’s Sonosite SII POC ultrasound system helps to keep crowded areas clearer with a small ultrasound footprint.

Feature | Ultrasound Imaging | July 07, 2020
With the miniaturization of technology, improved ease of use, lower system cost, increased portability and greater ac
A patient implanted with the Axonics System can undergo MRI examinations safely with radio frequency (RF) Transmit Body or Head Coil under the conditions outlined in the Axonics MRI Conditional Guidelines.

A patient implanted with the Axonics System can undergo MRI examinations safely with radio frequency (RF) Transmit Body or Head Coil under the conditions outlined in the Axonics MRI Conditional Guidelines.

News | Magnetic Resonance Imaging (MRI) | July 02, 2020
July 2, 2020 — Axonics Modulation Technologies, Inc., a medical technology company that has developed and is commerci
This data represents wave 2 of a QuickPoLL survey conducted in partnership with an imagePRO panel created by The MarkeTech Group (TMTG), regarding the effects of COVID-19 on their business

Getty Images

Feature | Coronavirus (COVID-19) | July 01, 2020 | By Melinda Taschetta-Millane
A 3-D ultrasound system provides an effective, noninvasive way to estimate blood flow that retains its accuracy across different equipment, operators and facilities, according to a study published in the journal Radiology.

Volume flow as a function of color flow gain (at a single testing site). For each row the color flow c-plane and the computed volume flow are shown as a function of color flow gain. The c-plane is shown for four representative gain levels, whereas the computed volume flow is shown for 12–17 steps across the available gain settings. Flow was computed with (solid circles on the graphs) and without (hollow circles on the graphs) partial volume correction. Partial volume correction accounts for pixels that are only partially inside the lumen. Therefore, high gain (ie, blooming) does not result in overestimation of flow. Systems 1 and 2 converge to true flow after the lumen is filled with color pixel. System 3 is nearly constant regarding gain and underestimates the flow by approximately 17%. Shown are mean flow estimated from 20 volumes, and the error bars show standard deviation. Image courtesy of the journal Radiology

News | Ultrasound Imaging | July 01, 2020
July 1, 2020 — A 3-D ultrasound
Researchers reviewed results of prostate biopsies on over 3,400 men who had targets identified on prostate MRI and found that the positive predictive value of the test for prostate cancer was highly variable at different sites
News | Prostate Cancer | July 01, 2020
July 1, 2020 — Prostate MRI is an emerging technology used to identify and guide treatment for...
R2* maps of healthy control participants and participants with Alzheimer disease. R2* maps are windowed between 10 and 50 sec21. Differences in iron concentration in basal ganglia are too small to allow visual separation between patients with Alzheimer disease and control participants, and iron levels strongly depend on anatomic structure and subject age. Image courtesy of Radiological Society of North America

R2* maps of healthy control participants and participants with Alzheimer disease. R2* maps are windowed between 10 and 50 sec21. Differences in iron concentration in basal ganglia are too small to allow visual separation between patients with Alzheimer disease and control participants, and iron levels strongly depend on anatomic structure and subject age. Image courtesy of Radiological Society of North America

News | Magnetic Resonance Imaging (MRI) | July 01, 2020
July 1, 2020 — Researchers using magnetic...
Imaging Artificial Intelligence (AI) provider Qure.ai announced its first US FDA 510(k) clearance for its head CT scan product qER. The US Food and Drug Administration's decision covers four critical abnormalities identified by Qure.ai's emergency room product.
News | Artificial Intelligence | June 30, 2020
June 30, 2020 — Imaging Artificial Intelligence (AI) provider Qure.ai announced its first US FDA 510(k) clearance for
In new QuickPoLL survey on imaging during the pandemic, responses were tallied from around 170 radiology administrators and business managers, who are part of an imagePRO panel created by The MarkeTech Group (TMTG), regarding the effects of COVID-19 on their business. TMTG is a research firm specializing in the medical device, healthcare and pharmaceutical industries.
Feature | Coronavirus (COVID-19) | June 30, 2020 | By Melinda Taschetta-Millane