News | Magnetic Resonance Imaging (MRI) | February 17, 2021

New study shows there is a lot to consider

T1 structural images for the two sequences, MPRAGE and MPRAGE+PMC. The top row shows the MPRAGE sequence, while the bottom row shows the images that were generated with the MPRAGE+PMC sequence. Columns represent two different participants, one with minimal head motion (left, Low-Mover) and another with a large quantity of motion (right, High-Mover). Pial and white matter (WM) surface reconstruction from Freesurfer are also shown.

T1 structural images for the two sequences, MPRAGE and MPRAGE+PMC. The top row shows the MPRAGE sequence, while the bottom row shows the images that were generated with the MPRAGE+PMC sequence. Columns represent two different participants, one with minimal head motion (left, Low-Mover) and another with a large quantity of motion (right, High-Mover). Pial and white matter (WM) surface reconstruction from Freesurfer are also shown.


February 17, 2021 — A new paper, "Is it time to switch your T1W sequence? Assessing the impact of prospective motion correction on the reliability and quality of structural imaging," evaluates two different ways of collecting magnetic resonance imaging (MRI) structural images. The first type of sequence, called MPRAGE, has been widely used by the neuroimaging research community for the last 20-plus years. More recently, an updated version of this sequence that includes Prospective Motion Correction (PMC) has been developed. The sequence with PMC adapts the MPRAGE sequence based on participant head motion. That makes it ideal for collecting data on children, who often struggle to hold still during MRIs. However, researchers are reluctant to change MRI sequences, since that typically means that they will not be able to use old and new data together in a statistical analysis.

"That's where this study comes in," said Alexandre Franco, Computational Neuroimaging Laboratories Director at The Nathan S. Kline Institute for Psychiatric Research and a research scientist at the Child Mind Institute. "It directly compares the reliability of the two sequences, MPRAGE and MPRAGE+PMC."

By using data collected for a Healthy Brain Network initiative that performed two MPRAGE and two MPRAGE+PMC sequences within the same imaging session, with one MPRAGE (MPRAGE1) and MPRAGE+PMC (MPRAGE+PMC1) pair at the beginning of the session and another pair (MPRAGE2 and MPRAGE+PMC2) at the end of the session, the new study directly compares two sequences from 348 participants. The study aimed to determine if the morphometric measurements obtained from both protocols are equivalent or if one sequence has a clear advantage over the other, while also comparing them through other quality control measurements. Results show that there is a high inter-sequence reliability between the MPRAGE and MPRAGE+PMC sequences. Results also indicate that the MPRAGE+PMC sequence is robust, but not impervious, to high head motion. "In the end use we suggest that studies targeting hyperkinetic populations use the MPRAGE+PMC sequence, given its higher robustness to head motion and higher reliability scores," said Franco.

For more information: www.childmind.org


Related Content

News | Magnetic Resonance Imaging (MRI)

April 27 2026 — SimonMed, one of the nation’s largest independent outpatient imaging providers, has announced the ...

Time May 04, 2026
arrow
News | Computed Tomography (CT)

April 23, 2026 — Royal Philips has received 510(k) clearance from the U.S. Food and Drug Administration (FDA) for its ...

Time April 30, 2026
arrow
News | X-Ray

April 29, 2026 — Results from a new study* presented at the American Roentgen Ray Society’s (ARRS) 2026 annual meeting ...

Time April 29, 2026
arrow
News | Radiology Business

April 28, 2026 — The American Society of Radiologic Technologists will award Life Member status to three longstanding ...

Time April 29, 2026
arrow
News | Imaging Software Development

April 28, 2026 — Avatar Medical has been granted FDA 510(k) clearance for Avatar Medical Vision, its software platform ...

Time April 28, 2026
arrow
News | Radiology Business

April 24, 2026 — The 2026 vacancy rate for radiation therapists decreased to 11.4% and the vacancy rate for medical ...

Time April 24, 2026
arrow
News | Contrast Agents

April 23, 2026 — On April 23, GE HealthCare announced the first patient has been dosed in the international, multi ...

Time April 23, 2026
arrow
News | Artificial Intelligence

April 20, 2026 — DeepTek, provider of the Augmento platform and deepc, the company behind deepcOS, have introduced a ...

Time April 23, 2026
arrow
News | Radiology Imaging

April 20, 2026 — Bracco Imaging has announced a strategic alliance with NYU Langone Health to advance innovation in ...

Time April 23, 2026
arrow
News | X-Ray

April 14, 2026 — KA Imaging is seeing continued adoption of its X-ray technology across new regions, with recent ...

Time April 15, 2026
arrow
Subscribe Now