News | Radiation Dose Management | January 03, 2017

Gonadal Shielding Effectively Minimizes Radiation Dose

Gonadal Shielding Minimizes Radiation Dose, Vivian Fornell

Gonadal shielding (triangular plate) of a toddler undergoing a hip X-ray to reduce the radiation dose to the patient.

January 3, 2017 — The use of a flat contact shield during imaging of the adult male pelvis can result in statistically significant radiation dose reduction to the testes, according to a recent study published in the journal Radiologic Technology.

Conducted by Terri L. Fauber, Ed.D., R.T.(R)(M), a professor emeritus of radiation sciences at Virginia Commonwealth University in Richmond, Virginia, the study shows a 36.4 percent increase in exposure to the testes was calculated when no contact shield was used during pelvic imaging. As a result, Fauber concludes the routine practice of shielding adult male gonads during radiographic imaging of the pelvis should be a best practice for all radiographers.

Fauber first completed a literature review to identify historical trends in shielding radiosensitive organs. The literature review showed a wide disparity of opinions on the use of gonadal shielding. “The effectiveness of gonadal shielding has been a debated topic in clinical practice for many years and it continues today with just as much zeal,” Fauber said. “I was surprised to find the literature also debated the efficacy of gonadal shielding. As a result, I was compelled to further explore the topic.”

To determine whether shielding reduces radiation exposure to the gonads, Fauber used a male anthropomorphic phantom equipped with holes for placement of thermoluminescent dosimeters to correspond to the testes. A flat contact shield was used to shield the phantom’s testes. Ten exposures were made of the pelvis with and without shielding. The exposed TLDs were then analyzed by the University of Wisconsin Calibration Laboratory, which reported the radiation exposure readings for the TLDs with and without shielding.

The study notes that the findings are at odds with previous research that concluded radiation exposure to the testes is not significantly reduced by the use of flat contact shielding. However, Fauber’s findings lend support to a 2010 study that showed that radiation exposure to the testes can be reduced significantly when using a flat contact shield.

“Because there are uncertainties regarding the biologic risks of low-dose radiation exposure, gonadal shields should be consistently utilized and accurately placed,” Fauber said. “In addition it’s good radiation hygiene, especially as our patients become more knowledgeable about potential risks of radiation exposure.”

Fauber said that very little research exists regarding the attitudes of radiologic technologists toward the importance of gonadal shielding and self-reported practices. A large study in the United States on self-reported knowledge about the efficacy of gonadal shielding and its routine practice would provide important information on shielding as a radiation safety practice. Investigating the effectiveness of staff training on the practice of correctly shielding the gonads also would provide valuable information to encourage radiographers to use gonadal shielding routinely.

The study "Gonadal Shielding in Radiography: A Best Practice?" was published in the American Society of Radiologic Technologist journal, Radiologic Technology, November/December 2016, Volume 88, Number 2. 

Related Content

Samsung GC70 Digital Radiography System Receives FDA 510(k) Clearance
Technology | Digital Radiography (DR) | April 27, 2017
Samsung announces that the GC70 Digital Radiography (DR) system received U.S. Food and Drug Administration (FDA) 510(k...
Colorado Hospital Upgrades X-ray Systems to Enhance Care for Residents, Tourists
News | Digital Radiography (DR) | April 20, 2017
Estes Park Medical Center (Estes Park, Colo.) recently upgraded its diagnostic imaging department with high-resolution...
Sponsored Content | Videos | Digital Radiography (DR) | April 20, 2017
Rob Fabrizio, director of strategic marketing at Fujifilm Medical Systems USA, discusses the latest innovations in di
University of California Study Searches for Consistent CT Dose Best Practices
News | Radiation Dose Management | April 17, 2017
A new study led by UC San Francisco has found that radiation doses can be safely and effectively reduced – and more...
JPI Healthcare Solutions Installs CareRay X-Vue DR System at Ortho Northwest
News | Digital Radiography (DR) | April 12, 2017
April 12, 2017 — JPI Healthcare has completed the installation of a CareRay dual panel X-Vue...
Bayer, Radimetrics Enterprise Platform, radiation dose management, Connecticut Hospital Association, statewide repository
News | Radiation Dose Management | March 30, 2017
The Connecticut Hospital Association (CHA) and Bayer announced an alliance to establish the United States' first-ever...
Medsquare, VirtualDose CT module, organ dose calculation, DACS Radiation Dose Monitor, RDM
Technology | Radiation Dose Management | March 24, 2017
Medsquare announced the integration of the organ dose module VirtualDose CT into its DACS (dose archive and...
Carestream, 2017 NFL Scouting Combine, digital X-ray, cone beam CT, CBCT
News | Digital Radiography (DR) | March 23, 2017
Carestream announced that it provided a DRX Plus 3543 digital X-ray detector for the 2017 National Football League (NFL...
Carestream, UnityPoint Health - Meriter, Wisconsin, DRX-Evolution Plus, digital X-ray, DR systems, orthopedic imaging
News | Digital Radiography (DR) | March 16, 2017
UnityPoint Health – Meriter (Madison, Wis.) recently installed two Carestream DRX-Evolution Plus systems as part of an...
Overlay Init