Fujifilm Medical Systems Removes Older Duodenoscopes From Clinical Use

January 16, 2017 — Fujifilm Medical Systems has informed the U.S. Food and Drug Administration (FDA) of its plans to remove its legacy 250/450 duodenoscope models from clinical use. This is based on the limited number currently in use and issues related to reprocessing of these devices leading to reports of multi‐drug resistant bacteria on endoscopes used for endoscopic retrograde cholangiopancreatogram (ERCP) procedures. 

On Jan. 13, 2017, Fuji notified its customers it will replace the 250/450 duodenoscope models with the newer ED-530XT model, in addition to necessary accessories (i.e. brushes) at no cost.

In December 2015, Fuji issued validated manual reprocessing instructions for the ED-530XT duodenoscope to replace those provided in the original device labeling. At that time, the FDA recommended that users and healthcare facilities apply the revised reprocessing instructions for the ED-530XT duodenoscope to reprocess the older 250/450 duodenoscope models while formal validation testing continued for these particular models.

Read letter sent to Fujifilm customers at www.fujifilmusa.com/products/medical/endoscopy/pdf/urgent_field_correction_ed_530XT.pdf

The validated manual reprocessing procedures for the ED-530XT duodenoscope outlined in a December 2015 FDA safety communication remain the same. Healthcare facilities should continue to use these validated instructions when reprocessing Fuji ED-530XT duodenoscope models. The validated reprocessing instructions when followed correctly, are intended to effectively clean and high-level disinfect the Fuji ED-530XT duodenoscope.

The FDA recommendations for healthcare facilities and staff include:
  • Remove from circulation and return your facility’s 250/450 duodenoscope models to Fuji as outlined in Fuji’s Customer Notification Letter.
  • Train appropriate staff on Fuji’s validated reprocessing instructions for the ED-530XT model and implement them as soon as possible.
  • Implement the reprocessing procedures for Fuji’s ED-530XT duodenoscope in accordance with the manufacturer’s reprocessing instructions issued in December 2015.
  • Immediately remove from service to assess, repair and/or replace any duodenoscope that shows signs of damage. Examples of damage may include: loose parts, damaged channel walls, kinks or bends in tubing, holes in the distal end, or other signs of wear or damage.
 • Contact your Fuji sales representative if you have any questions or concerns regarding this removal, a damaged device, or the validated reprocessing instructions for the ED-530XT duodenoscope.

The FDA said it continues to closely monitor the association between reprocessed endoscopes and the transmission of infectious agents by working closely with duodenoscope and automated endoscope reprocessor manufacturers as they validate their reprocessing instructions to a level adequate to clean and high-level disinfect their devices. The FDA also is communicating recommendations to healthcare providers and end users to help mitigate the risk associated with infection transmission and reusable medical devices.

Visit the Infections Associated with Reprocessed Duodenoscopes webpage for a complete listing of actions the agency has taken on this issue. The FDA said it will continue to provide updates as appropriate.


Related Content

News | Ultrasound Imaging

Aug. 14, 2025 — Philips has announced a plan for new investments of more than $150 million in U.S. manufacturing and ...

Time August 14, 2025
arrow
News | RSNA

Aug. 13, 2025 — Registration is now open for the RSNA 111th Scientific Assembly and Annual Meeting, the world’s leading ...

Time August 13, 2025
arrow
News | Remote Viewing Systems

August 7, 2025 — DeepHealth, Inc., a wholly-owned subsidiary of RadNet, Inc., announced it has received FDA 510(k) ...

Time August 07, 2025
arrow
Feature | X-Ray | By Siân Phillips, MD

Imaging is a crucial diagnostic, treatment and monitoring tool in modern medicine. From diagnosis to disease management ...

Time July 31, 2025
arrow
News | Digital Radiography (DR)

July 30, 2025 — Fujifilm Healthcare Americas Corp., a leading provider of diagnostic and enterprise imaging solutions ...

Time July 30, 2025
arrow
News | Advanced Visualization

July 28, 2025 — Frost & Sullivan has named Siemens Healthineers the 2025 North America Company of the Year in the ...

Time July 28, 2025
arrow
News | X-Ray

July 24, 2025 — GE HealthCare has announced the commercial availability of an advanced floor-mounted digital X-ray ...

Time July 24, 2025
arrow
News | Artificial Intelligence

July 17, 2025 — HOPPR, a secure AI development platform for medical imaging, has announced the commercial release of its ...

Time July 17, 2025
arrow
News | Breast Imaging

July 17, 2025 — RadNet, Inc., a provider of high-quality, cost-effective diagnostic imaging services and digital health ...

Time July 17, 2025
arrow
News | Focused Ultrasound Therapy

July 15, 2025 — Newswise —The UK Focused Ultrasound Foundation and Pancreatic Cancer UK have announced a new partnership ...

Time July 16, 2025
arrow
Subscribe Now