RSNA 2012 Mammography Thyroid Shields

December 7, 2012 — The radiation dose to areas of the body near the breast during mammography is negligible, or very low, and does not result in an increased risk of cancer, according to a clinical study presented at the annual meeting of the Radiological Society of North America (RSNA). The results suggest that the use of thyroid shields during mammography is unnecessary.

"Thyroid shields can impede good mammographic quality and, therefore, are not recommended during mammography," said Alison L. Chetlen, D.O., assistant professor of radiology at Penn State Hershey Medical Center.

During mammography, some X-rays scatter away from the primary beam in the breast and spread outward in different directions. Although this scatter radiation is much weaker than the primary beam, there has been concern that women exposed to it during mammography could face an increased risk of cancer, especially in radiosensitive areas like the thyroid gland.

To better understand the potential impact of scatter radiation, Chetlen and colleagues set out to measure the dose received by the thyroid gland, salivary gland, sternum, uterus and the lens of the eye during screening digital mammography. Each of the 207 women in the study group wore six optically stimulated luminescent dosimeters — a device used to measure an absorbed dose of ionizing radiation — while undergoing two-view screening mammography.

Analysis of the dosimeters by a medical physicist immediately after the exam revealed that the doses to the various areas outside of the breast ranged from negligible to very low.

Absorbed radiation dose is measured in a unit called a milligray (mGy). The average estimated organ dose to the salivary gland was 0.05 mGy. The average estimated organ dose to the thyroid gland was 0.05 mGy. These doses are only a fraction of the radiation people are exposed to from natural background sources, such as cosmic radiation and radionuclides in the ground. In fact, all areas except for the sternum received less than 2 percent of annual background radiation dose.

Measured dose to the bridge of the eye and umbilicus was negligible, indicating no increased risk to the patient of cataracts or interference with normal embryonic development in early pregnancy.

"The risk of cancer induction at these low levels is indistinguishable from background incidence of cancer due to other sources," Chetlen said.

The findings are particularly important in light of a recent increase in the incidence of thyroid cancer, one of the most radiosensitive of all cancers. The number of thyroid cancer diagnoses in women nearly doubled from 2000 to 2008, leading some to suspect that mammography may be a contributing factor and that women should wear lead thyroid shields during exams, an idea that Chetlen and other mammography experts strongly discourage.

Based on the extremely low scatter radiation dose to the thyroid — equivalent to just a few minutes of background radiation — thyroid shields are unnecessary during mammography. In addition, the researchers warn that use of thyroid shields could result in an increased radiation dose to patients.

"A thyroid shield gets in the way of the exam and can actually cause an increase in radiation dose by necessitating repeat exams," Chetlen said.

Chetlen also pointed out that the thyroid gland is far less radiosensitive after age 30. The American Cancer Society and other organizations recommend that women have mammography screening once every year, beginning at age 40.

"In the age group eligible for screening, the thyroid gland is not very radiosensitive," Chetlen said.

Coauthors are Steven King, M.S., Karen Brown, C.H.P., D.A.B.R., Brian Lorah, Susann Schetter, D.O., Claudia Kasales, M.D., Shelley Tuzzato, R.T.R.M., and Shelly Rambler, R.T.R.M.

For more information: www.rsna.org


Related Content

News | Mammography

April 29, 2025 — iCAD, a global provider of clinically proven AI-powered cancer detection solutions, has announced a ...

Time April 29, 2025
arrow
News | Mammography

April 24, 2025 — GE HealthCare will feature its latest advancements in diagnostic accuracy and patient-centered breast ...

Time April 24, 2025
arrow
News | Artificial Intelligence

March 10, 2025 — Lunit, a provider of AI-powered solutions for cancer diagnostics and therapeutics, has published a ...

Time March 10, 2025
arrow
News | Artificial Intelligence

Feb. 19, 2025 — SimonMed Imaging and HeartLung Technologies have signed a strategic partnership to offer HeartLung's AI ...

Time March 04, 2025
arrow
News | Ultrasound Imaging

Jan. 28, 2025 — GE HealthCare recently announced it has received 510(k) clearance from the United States Food and Drug ...

Time January 29, 2025
arrow
News | Breast Imaging

Jan. 8, 2025 — ScreenPoint Medical has acquiredf Biomediq A/S, a research-based company focused on the research ...

Time January 10, 2025
arrow
News | Breast Imaging

Dec.11, 2024 — iCAD, Inc., a provider of clinically proven AI-powered cancer detection solutions, recently announced ...

Time December 18, 2024
arrow
News | Mammography

Dec. 5, 2024 — At RSNA 2024, Lunit and Volpara Health announced their unified vision, focusing on a comprehensive ...

Time December 05, 2024
arrow
News | Mammography

Nov. 26, 2024 — GE HealthCare has introduced the Pristina Via* mammography system designed to enhance the screening ...

Time December 02, 2024
arrow
News | Artificial Intelligence

Dec. 2, 2024 — Lunit recently unveiled follow-up findings from the ScreenTrustCAD trial. This study, conducted at Capio ...

Time December 02, 2024
arrow
Subscribe Now