Biphasic Injection Protocol – Optimized for Coronary Artery Imaging

Stock image

This case is a routine coronary CTA with the focus on the coronary arteries. The patient was a middle-aged male with a family history of coronary artery disease, moderate risk profile and atypical chest pain. The EKG did not indicate myocardial ischemia and there was no enzyme leak. The referring physician was considering invasive angiography to definitely exclude coronary artery disease as the cause of the atypical chest pain. Due to concerns regarding the invasiveness of the procedure and a relatively low likelihood that coronary artery disease was present, a cardiac gated 64-slice coronary MDCT angiogram was performed instead.
The contrast injection protocol in this study is tailored to opacify the coronary arteries during the MDCT acquisition and to wash out the right atrium and ventricle with a saline chaser bolus.
The latter is desired for two reasons. First, if contrast were to remain in the right heart during acquisition, streak artifact may result in image degradation, potentially decreasing the evaluability of the mid-RCA. The streak artifact occurs because of the mixing of relatively dense contrast containing blood from the superior vena cava with in-flowing, non-opacified blood from the inferior vena cava. Secondly, if contrast material is located in the right heart during the scan acquisition, it obviously cannot contribute to the opacification of the coronary arteries — the structures of interest. A higher contrast volume at a higher rate, therefore, would be necessary to achieve the same opacification as in a patient with washed out right heart and optimal contrast timing.
Because the image quality in this case is excellent and the negative predictive value of coronary CTA in research studies is typically between 95 and 100 percent, it could be established that in this patient the chest symptoms were unrelated to coronary artery disease. The patient was spared the invasive angiogram.

Related Content

Two brain metastases from primary lung cancer are contrast enhanced in the brain of a 61-year-old male. Speakers at AHRA 2019 will state that ProHance and other macrocyclic MR agents present a very low risk to patients. Images courtesy of Bracco

Two brain metastases from primary lung cancer are contrast enhanced in the brain of a 61-year-old male. Speakers at AHRA 2019 will state that ProHance and other macrocyclic MR agents present a very low risk to patients. Images courtesy of Bracco

Feature | Contrast Media | July 18, 2019 | By Greg Freiherr
Macrocyclic contrast agents have the best safety profile of all the magnetic resonance (MR) contrast media that are n
FDA Approves Bayer's Gadavist Contrast for Cardiac MRI in Adult Coronary Artery Disease Patients
Technology | Contrast Media | July 15, 2019
The U.S. Food and Drug Administration (FDA) has approved Gadavist injection for use in cardiac magnetic resonance...
ITN Wins Jesse H. Neal Award for Best Technical Content
News | Radiology Business | April 24, 2019
April 24, 2019 — Imaging Technology News (ITN) was recently named the 2019 Jesse H.
Example of full-dose, 10 percent low-dose and algorithm-enhanced low-dose. Image courtesy of Enhao Gong, Ph.D.

Example of full-dose, 10 percent low-dose and algorithm-enhanced low-dose. Image courtesy of Enhao Gong, Ph.D.

Feature | Contrast Media Injectors | April 11, 2019 | By Jeff Zagoudis
One of the most controversial issues in radiology in recent years has been the use of...
Non-Contrast MRI Effective in Monitoring MS Patients
News | Neuro Imaging | March 18, 2019
Brain magnetic resonance imaging (MRI) without contrast agent is just as effective as the contrast-enhanced approach...
Ultrasound Societies Urge FDA to Remove "Black Box" on Ultrasound Contrast Agents
News | Ultrasound Imaging | March 07, 2019
National and international ultrasound societies are urging the U.S. Food and Drug Administration to remove the “black...
Sponsored Content | Case Study | Contrast Media | January 30, 2019
The medical imaging market relies heavily on contrast media, injected into patients to increase the contrast of bodil
Sponsored Content | Videos | Contrast Media | January 22, 2019
ITN Contributing Editor Greg Freiherr speaks with Guerb
Videos | Artificial Intelligence | December 21, 2018
Enhao Gong, Ph.D., founder of Subtle Medical, an artificial intelligence (AI) company that develops products to help
Guerbet Showcases Diagnostic and Interventional Imaging Solutions at RSNA 2018
News | Interventional Radiology | December 12, 2018
Guerbet LLC USA highlighted new and next-level product offerings and partnerships in contrast media, injectors,...