Case Study | October 02, 2008 | Suhny Abbara, M.D., assistant professor, Harvard Medical School, director, Cardiovascular Imaging, M
Stock image

This case is a routine coronary CTA with the focus on the coronary arteries. The patient was a middle-aged male with a family history of coronary artery disease, moderate risk profile and atypical chest pain. The EKG did not indicate myocardial ischemia and there was no enzyme leak. The referring physician was considering invasive angiography to definitely exclude coronary artery disease as the cause of the atypical chest pain. Due to concerns regarding the invasiveness of the procedure and a relatively low likelihood that coronary artery disease was present, a cardiac gated 64-slice coronary MDCT angiogram was performed instead.
The contrast injection protocol in this study is tailored to opacify the coronary arteries during the MDCT acquisition and to wash out the right atrium and ventricle with a saline chaser bolus.
The latter is desired for two reasons. First, if contrast were to remain in the right heart during acquisition, streak artifact may result in image degradation, potentially decreasing the evaluability of the mid-RCA. The streak artifact occurs because of the mixing of relatively dense contrast containing blood from the superior vena cava with in-flowing, non-opacified blood from the inferior vena cava. Secondly, if contrast material is located in the right heart during the scan acquisition, it obviously cannot contribute to the opacification of the coronary arteries — the structures of interest. A higher contrast volume at a higher rate, therefore, would be necessary to achieve the same opacification as in a patient with washed out right heart and optimal contrast timing.
Because the image quality in this case is excellent and the negative predictive value of coronary CTA in research studies is typically between 95 and 100 percent, it could be established that in this patient the chest symptoms were unrelated to coronary artery disease. The patient was spared the invasive angiogram.

Related Content

News | Contrast Media

August 3, 2022 — According to an open-access article in ARRS’ American Journal of Roentgenology (AJR), electronic health ...

Time August 03, 2022
arrow
News | Contrast Media

August 2, 2022 – A new update has been announced as the radiology world continues to address supply disruptions of ...

Time August 02, 2022
arrow
News | Contrast Media

July 18, 2022 — Fresenius Kabi announced today it will introduce a portfolio of generic contrast media agents in the ...

Time July 18, 2022
arrow
Sponsored Content | Case Study | SPECT-CT

A patient’s chance of recovery can be directly impacted when a diagnosis is missed or delayed. To get the right ...

Time July 06, 2022
arrow
Feature | Radiology Business | By Melinda Taschetta-Millane

Here is what you and your colleagues found to be most interesting in the field of medical imaging during the month of ...

Time July 05, 2022
arrow
News | Contrast Media

June 2, 2022 — Today, GE Healthcare issued a statement that it expects production capacity from its Shanghai facility to ...

Time June 02, 2022
arrow
Feature | Radiology Business | By Melinda Taschetta-Millane

Here is what you and your colleagues found to be most interesting in the field of medical imaging during the month of ...

Time June 01, 2022
arrow
News | Contrast Media

June 1, 2022 — The current shortage of iodinated contrast in the United States due to the COVID-19 related production ...

Time June 01, 2022
arrow
News | Artificial Intelligence

May 31, 2022 — Imaging Biometrics, LLC (IB), a subsidiary of IQ-AI Limited, announced it has submitted an FDA 510(k) ...

Time May 31, 2022
arrow
News | Contrast Media

May 19 2022 — Recent disruptions in a pharmaceutical supply chain have impacted the global availability of GE Healthcare ...

Time May 19, 2022
arrow
Subscribe Now